Automatic Hematology Analyzer KT-HA3100
SKU: MW-KT-HA3100
Micro detection Touch operation Automatic storage Environmentally friendly The Automatic Hematology Analyzer is an in vitro diagnostic device that used for quantitative analysis of blood cells, and it can realize three classifications of the count results of white blood cells. This analyzer is a clinical inspection instrument for screening.
Price on request
Sign in and submit a request to see pricing for this product.
Specifications
| Model | KT-HA3100 |
| Display | 8" color LCD touch screen; |
| Accuracy | WBC ≤ ±6.0%, RBC ≤ ±5.0%, HGB ≤ ±4.0%, PLT ≤ ±10.0%, HCT ≤ ±5.0%; |
| Interface | USB interface, RS232 interface; |
| Precision | WBC ≤ 3.0%, RBC ≤ 2.0%, HGB ≤ 2.0%, PLT ≤ 5.0%, MCV ≤ 0.5%; |
| Carry-over | WBC ≤ 2.0%, RBC ≤ 1.0%, HGB ≤ 2.0%, PLT ≤ 2.0%; |
| Test speed | 60 specimens/hour, continuous working for 24 hours; |
| Blank count | WBC ≤ 0.2×10⁹/L, RBC ≤ 0.02×10¹²/L, HGB ≤ 1 g/L, PLT ≤ 10×10⁹/L, HCT ≤ ±0.5%; |
| Data storage | It can automatically store more than 300,000 groups of complete results (each result has three histograms); |
| Power supply | ~100V-240V; 50/60Hz. |
| Sample volume | trace detection, 10μL peripheral blood or anti-coagulant blood, pre-dilution mode 20μL; |
| Test parameters | WBC 3-part differential counting, 24 parameters (including WBC, RBC, PLT color histogram); |
| Country of Origin | China |
| Detection channel | dual channel; |
| Linearity deviation | WBC ≤ ±5%, RBC ≤ ±5%, HGB ≤ ±3%, PLT ≤ ±10%; |
| Measurement principle | counting by electrical impedance method, colorimetric method to measure HGB; |
| Quality control method | L-J, automatic drawing and printing of quality control charts; |
| Linearity related coefficient | WBC ≥ 0.990, RBC ≥ 0.990, HGB ≥ 0.990, PLT ≥ 0.990; |