1 / 5
Infusion Pump HK-T2000
SKU: MW-HK-T2000
Price on request
Sign in and submit a request to see pricing for this product.
Specifications
| Size | 97x 185x 126 (L*H*W)mm |
| VTBI | (0.00~99999.99)ml,increment at0.01ml |
| Alarm | Use battery, Door open, Occusion, Almost done, No operation, Infusion completion, LowBattery, Battery exhausted, Malfunction etc. |
| Model | HK-T2000 |
| Battery | Li-Polymer 7.2V 6800mAh;12hoursat25ml/h |
| Accuracy | ±5% |
| Features | Anti-Bolus,Drug library,Anti-reverse detection,Double CPU,Touch screen, Screen lock Screen brightness adjustable, Key blacklight adjustable,Bolus,Drop sensor |
| KVO rate | (0.10~10.00)ml/h, incrementat0.01ml/h |
| Drip rate | (1-400)d/min, increment at ld/min |
| Netweight | 1.62KG |
| BOLUS rate | (0.10~2001.00)ml/h, increment at 0.01ml/h |
| Bolus VTBI | (0.00~100.00) ml, increment at0.01ml |
| Night mode | Available,Screen brightness adjustable |
| Purge rate | (0.10~2001.00)ml/h, increment at 0.01ml/h |
| Waterproof | IP24 |
| Power supply | AC:100-240V 50/60Hz;DC:12V±1.2V |
| Infusion Mode | Rate Mode. Drip Mode, Time Mode,Weight Mode, Dose Mode,Intermittent Mode,TPN Mode,Shift Mode, Program Mode,Micro Mode, Multi Mode,DrugLibrary |
| Infusion rate | (0.10~2001.00)ml/h, increment at 0.01ml/h |
| Classification | Class 1,Type CF |
| History record | 30000 |
| Package Weight | 2.5KG |
| Volume infused | (0.00~99999.99)ml |
| Bubble detection | 6 levels (25μl-800 μl) |
| Power consumption | 35VA |
| Occlusion pressure | 13 levels(10~160)kPa |
| Optional functions | Nurse call,WIFI,Heating, Dedicated tube |
| Package Dimensions | 29*22*21cm |
| Atmosphere pressure | 50.0kPa~106.0kPa |
| Working environment | Temperature:-20*C~+60C;Relative humidity:10%-95% |
| Applicable infusion set | Various brands of standard infusion sets |